Faults And Error Codes - IKA ULTRA-TURRAX T 65 basic Instrucciones De Uso

Tabla de contenido
Faults on the T 65 basic ULTRA-TURRAX
Cause
Power supply failure
Stand not in position
(if the stand safety limit
switch is connected)
Overload
If none of these causes was the cause of the problem, a fault other than those described must have occurred and the
unit must be checked by an authorized person.
Error codes T 65 digital ULTRA-TURRAX
In the event of a malfunction during operation, proceed as follows:
- Switch off the device at the power switch (A).
- Carry out corrective measures.
- Restart the device.
All error codes are displayed with the display "trip" i.e. it appears alternately "trip" and the corresponding error code.
Error code
Cause
lt.AC
Motor overload
Oht1
Overtemperature
OI.AC
Overcurrent on the
frequency inverter
OU
Overvoltage at the
frequency inverter
Note: High-frequency noises during operation of the T 65 digital ULTRA-TURRAX
cy inverter, depending on the design. They certainly do not indicate a defect within the unit.
Note: The green and red LEDs around the "Speed setting" rotary knob (E) are not relevant to the T 65 digital ULTRA-TURRAX
Any faults are shown exclusively on the display (F).
Note: If any error codes other than those listed above appear on the display please switch the unit off and on again at the main
switch (A). If the error code persists please contact the IKA Service Department.
If the actions described fail to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the IKA service department
- Send the device for repair, including a short description of the fault.
20

Faults and error codes

®
Effect
The unit cannot be switched on at the main
switch.
The equipment has switched itself off.
The unit cannot be switched on at the main
switch.
The equipment has switched itself off.
The over-current limit switch has switched
the unit off.
®
Effect
Solution
Device off
Reduce the load by reducing the speed.
Overloads are especially liable to occur with high-viscosity media and at
high speeds.
Device off
Switch the unit off and allow it to cool down, try reducing the load or the
operating cycle.
Device off
Reduce the load by reducing the speed.
If this error appears even when idling with no load, the frequency inverter
is defective. Consult the Service Engineer.
Device off
Check the mains power cable values and the power supply.
Solution
Checking the power supply should be performed by
authorized personnel.
Precautionary measure:
Check the power supply
Move the stand into position.
Precautionary measure:
Check the power supply
Remedy the overload, e.g.:
- The viscosity of medium is too high
- Solids jamming the stator / rotor
- Dispersing element defective (blocked).
may to some extent originate from the frequen-
®
unit.
®