Nebulizer Use instructions
Disinfecting
All nebuliser components (except air hose) can be disinfected with chemical disinfectants following dosage and use limitations
as provided by disinfec tant manufacturer. Disinfectants are usually available at pharmacies.
Sterilizing with steam
All nebuliser components (except air hose and ma-sks) can be heat steam sterilized up to 121 °C (20 min.) or 134 °C (7 min.).
EN554/ISO11134.The sterilization packaging must conform to EN868/ISO11607 and be suitable for steam sterilization.
After sterilization always let all components cool down to ambient temperature before further use.
Do not repeat sterilization cycle when compo-nents are still warm.
MAINTENANCE, CARE, AND SERVICE
Order all spare parts from your dealer or pharmacis.
It is advisable replace nebulizer after some 100 to 120 treatments on single patient or after about 20 sterilization cycles. Check
the filter continually for cleanliness and repla-ce it if dirty, or after a maximum of 3 months use. Spare filters are provided with
the device. To replace the filter, open the air filter compartment 4 marked «Filter» and change the filter 11
MALFUNCTIONS AND ACTIONS TO TAKE
The device cannot be switched on
Ensure the power cord 2 is correctly plugged into the socket. Ensure the ON/OFF switch 3 is in the position «I». The device
has automatically switched off due to overheating. Wait until it has cooled down and try again.
The nebuliser functions poorly or not at all
Ensure the air hose 6 is correctly connected at both ends. Ensure the air hose is not squashed, bent, dirty or blocked. If necessa-
ry, replace with a new one. Ensure the nebuliser 5 is correctly assembled 12. Ensure the required medication has been added.

TECHNICAL DATA

Piston electrocompressor with thermal cut
Nominal voltage: see data label
Dimensions: 130x190x100 mm
Weight: 1.1 Kg approx.
Temporary use: 20/40 min I/O
MDD 93/42/EEC - Class of risk IIa
Maximum pressure: 185 kPa
Air flow: 10 Lt/min
Noise level: 58 dBA (EN13544-1)
Neb rate: 0.35 ml/min (con soluzione NaCl0.9%)
Operating pressure: 60 kPa (0.6bar)
Operating flow: 5 l/min
MMAD: 2.61 µm GSD: 2.75 Aerosol Output: 143 µl Output Rate: 87 µl/min
For any improvement reasons, Beper reserves the right to modify or improve the product without any notice.
The European directive (Dir. 2012/19/Ue-WEEE) on Waste Electrical and Elec-
tronic Equipment (WEEE), requires that old household electrical appliances must
not be disposed of in the normal unsorted municipal waste stream. Old appliances
must be collected separately in order to optimize the recovery and recycling of
the materials they contain, and reduce the impact on human health and the en-
vironment. The crossed out "wheeled bin" symbol on the product reminds you
of your obligation, that when you dispose of the appliance, it must be separately