FUSE REPLACEMENT
1. Turn the unit off and remove the power cord from both the wall outlet and the
Welch Allyn Light Source.
Fuse Holder
2. Use a small screwdriver to remove the fuse holder above the power cord
connector by releasing the spring-loaded snap connections.
3. Pull out the fuse holder.
4. Remove and replace with new fuses of the same type. There is no required
orientation of the fuses. Be sure to replace both fuses.
Snap
Connections
Fuses
5. Reinsert the fuse holder by pressing it into the power connector until it snaps into
place.
6. If the new fuse blows, immediately contact Welch Allyn Technical Service at 315-
685-4560.
Fuse Holder