Connection tubing
1. Preparation
Check the product:
– Expiry date of the cleaning agent/disinfectant
– Processing limits reached?
2. Cleaning and disinfection
Manual
cleaning:
Manual
disinfection:
Mechanical
cleaning with
disinfection:
3. Steam sterilisation
not usable
4. Visual inspection & storage
Inspect all individual parts. Replace any broken, misshapen or
seriously discoloured parts.
Storage location:
– dry
– dust-free
– protected from sources of contamination
optional: Use sterile packaging
not applicable
not applicable
Alkaline cleaning agent, e.g., neodisher
MediClean forte (Dr. Weigert) in conjunction with
neutralising agent, e.g., neodisher
Equipment:
– Cleaning device and disinfector in conformance
with DIN EN ISO 15883, e.g., RDG G7836 CD
(Miele)
– Special baskets for Miele instrument dishwasher
– Compressed air source for blowing dry
Vario TD programme or comparable valid
programmes
– 44 –
®
®
Z (Dr. Weigert)