Sensitive measuring devices must be checked from time to time for accuracy and leakage. We recommend
that the performance should be checked every 2 years.

GUIDANCE AND MANUFACTURE'S DECLARATION

The Digital Blood Pressure Monitor is intended for use in the electromagnetic environment specified below. The
customer of the user of the Digital Blood Pressure Monitor should assure that it is used in such an environment.
Emission test
Radio frequency (RF) emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
The Digital Blood Pressure Monitor is intended for use in the electromagnetic environment specified below. The
customer or the user of Digital Blood Pressure Monitor should assure that it is used in such an environment.
IEC 60601
Immunity test
Test level
Electrostatic
± 6 kV
discharge (ESD)
contact
IEC 61000-4-2
± 8 kV air
Power frequency
(50Hz/60Hz)
3 A/m
magnetic field IEC
61000-4-8
3 A/m
Radiated RF
80 MHz a
IEC 61000-4-3
2.5 GHz

ELECTROMAGNETIC EMISSION

Compliance
The Digital Blood Pressure Monitor use RF energy only for
its internal function. Therefore, its RF emissions are very
Group 1
low and are not likely to cause any interference in nearby
electronic equipment.
Class B
The Digital Blood Pressure Monitor is suitable for use in all
Not applicable
establishments other than domestic and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Not applicable

ELECTROMAGNETIC IMMUNITY

Compliance
level
± 6 kV
Floors should be wood, concrete or ceramic tile. If floor are covered
contact
with synthetic material, the relative humidity should be at least 30%.
± 8 kV aie
Power frequency magnetic fields should be at levels characteristic of
3 A/m
a typical location in a typical commercial or hospital environment.
Portable and mobile RF communications equipment should be used
no closer to any part of the Digital Blood Pressure Monitor, including
cables, than the recommended separation distance calculated from
the equation applicable to the frequency of the transmitter.
Recommended separation distance
d=1.167
P 80 MHz to 800 MHz
d=2.333
P 800 MHz to 2.5 GHz
P
Where
is the maximum output power rating of the transmitter in
3 A/m
watts (W) according to the transmitter manufacturer and
recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey
level in each frequency range
Interference may occur in the vicinity of
equipment marked with the following symbol:
Electromagnetic environment – guidance
Electromagnetic environment - guidance
A
, should be less than the compliance
B
d
is the
38