1. Choose Program number.
2. Start the run.
A
SET:
P2
2:00 h
600 V
50 mA
30 W
3. During a run you can view the
settings. It is also possible to interupt
the run and make changes in the
program.
4. Stop the run manually.
5. The run is stopped automatically.
Disconnect the leads and proceed with post-electrophoretic techniques. Since diffusion will begin as soon
as the voltage is turned off, you should remove the gel and begin staining, blotting or autoradiography
immediately.
5.4 Optional programming
You can disable the start current check that otherwise checks that the resistance is not higher than a
specified limit. This disabling function is set in the MORE mode instead of the SET mode in order not to
confuse users who do not need this feature, and for safety reasons.
The MORE mode is described in Fig. 5. After pressing MORE , the program asks you for the program
number and if you want to keep the start current check. Choose program number and change between
"YES" and "NO" by
that the function is disabled and high voltage can be applied despite a very high resistance. Leave the
MORE mode by EXIT.
w
Warning! By disabling the start current check, the power supply can deliver high voltage even if it is
not connected to electrophoresis equipment.
SET:
P1
200 V
40 mA
Select by
SET:
P2
600 V
50 mA
Start by
RUN:
P2
300 V
50 mA
set
RUN:
P2
500 V
50 mA
exit
or wait 5 s
RUN:
P2
600 V
50 mA
END:
P2
600 V
35 mA
Fig. 4 . Step-by-step summary of run and pause function
/
and confirm with ENTER. Choosing "NO" for start current check means
OFF
10 W
A
2:00 h
30 W
run
A
0:01 h
25 W
pause
continue
A
0:30 h
PAUSE:
25 W
500 V
pause
run
or
continue
A
1:45 h
END:
24 W
600 V
STOP
A
2:00 h
21 W
EPS 601 - User Manual
A
P2
0:30 h
50 mA
100 W
set
pause
or
exit
continue
A
P-SET:
P2
2:00 h
600 V
50 mA
30 W
Programming
A
P2
1:45 h
40 mA
24 W