Setting the suction level
4
Figure 4
1.
Clamp the connective tubing to
occlude the patient port
Clamp
5
Figure 5
2.
Turn the mode selector switch to | (On).
6
Figure 6
3.
Rotate the suction control knob until the
vacuum gauge indicates the required setting.
WARNING
⚠ The regulator must be occluded when setting the prescribed
Pre-Use Checkout Procedure
WARNING
⚠ The Pre-Use Checkout Procedure must be performed before
Important: All tests must be performed with supply vacuum of 40.0 kPa
(300 mm Hg) minimum.
1.
Turn the mode selector switch to O (Off).
2.
Rotate the suction control knob one full turn clockwise (increase).
3.
Clamp the connective tubing to occlude the fitting port. The gauge needle
should not move.
6
6700-0358-000 (Rev. 9) 08/2020
suction level so that the patient does not receive higher
than required suction.
using the equipment on each patient. If the regulator fails
any part of the Pre-Use Checkout Procedure, it must be
removed from service and repaired by qualified service
personnel.