Customer Contact Information; Prevena Plus™ Therapy System Specifications; Environmental Conditions - KCI PREVENA PLUS DUO Manual Del Usuario

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

CUSTOMER CONTACT INFORMATION

For questions regarding this product, supplies, maintenance or additional information about KCI
products and services, please contact KCI or a KCI authorized representative, or:
In the US call 1-800-275-4524 or visit www.acelity.com
Outside the US visit www.kci-medical.com
KCI USA, Inc. 12930 IH 10 West, San Antonio, TX 78249
PREVENA PLUS™ THERAPY SYSTEM SPECIFICATIONS
Dimensions: ....................................................................................... 3.5" W x 6.4" H x 2.16" D (8.9 x 16.3 x 5.49 cm)
Weight (with empty canister attached): .........................................................................................~.64 lbs (~.29 kg)
Pressure: ........................................................................................................................................... 125 mmHg (16.7 kPa)
Canister Volume: ........................................................................................................................................................ 150 mL
Electrical:
Battery Run Life: ...................................................................................................................................................~8.5 hours
Battery Charge Time: ................................................................................... ~6 hours from a fully discharged state
External Power Supply Input: ............................................................................. 100-240VAC 0.5A-0.3A 50 - 60 Hz
External Power Supply Output: ..........................................................................................................................5V, 2.0 A
Patient and Enclosure Leakage Current: ........................................................................................ <100 Microamps

Environmental Conditions:

Storage/Transport Conditions
Temperature Range: ............................................................................................................ 0°F (-18°C) to 140°F (60°C)
Relative Humidity Range: ...................................................................................................... 0-95% non-condensing
Operating Conditions
Temperature Range: ..............................................................................................................41°F (5°C) to 104°F (40°C)
Relative Humidity Range: .................................................................................................. 15 - 93% non-condensing
Atmospheric Pressure: .........................................1060 hpa (-1253 ft/-381. 9 m) to 700 hpa (9878 ft/3010 m)
Expected Service Life ..............................................................................................................................................7.5 days
IEC Classification
Medical Equipment
Type BF, Applied Part
Class II
IP22 - Protection against solid objects greater than 12.5 mm and against liquid water falling for short
periods of time.
The dressing components of the PREVENA PLUS DUO™ Incision Management System are considered
Applied Parts under IEC 60601-1.
Conforms to AAMI ES60601-1:2005 +A1:2012, IEC 60601-1-6 Ed.3.1:2013, IEC 60601-1-8 Ed.2.1:2012,
IEC 60601-1-11 Ed.2.0:2015, CSA C22.2#60601-1:2014 Ed.3