2 Intended use
2.1 Indications for use
The product is intended exclusively for lower limb exoprosthetic fittings.
2.2 Area of application
•
The maximum approved body weight is specified in the technical data (see Page 16).
2.3 Environmental conditions
Allowable environmental conditions
Temperature range for use: -10 °C to +60 °C
Storage and transport conditions: -10 °C to +60 °C, relative humidity: no restrictions
Moisture: fresh water, salt water, chlorinated water, sweat, urine
Dust, sand, highly hygroscopic particles (e.g. talcum)
Clean the product after it has come into contact with moisture/sand/dirt.
Unallowable environmental conditions
Acids
2.4 Service life
Prosthetic foot
An exact service life cannot be determined when the product is being used for fitness activities
since a very wide range of loads occur depending on the activity and intensity.
Heel wedge, forefoot pad
The product is a wear part, which means it is subject to normal wear and tear.
3 Safety
3.1 Explanation of warning symbols
Warning regarding possible risks of accident or injury.
CAUTION
Warning regarding possible technical damage.
NOTICE
3.2 General safety instructions
CAUTION
Usage under extreme loads
Risk of injury due to excessive loads during extreme activity
Although the product was developed for fitness activities, it may be damaged if exposed to
►
extreme loads.
Inspect the product for damage before and after each activity.
►
If damage is apparent or in case of doubt, do not continue using the product.
►
CAUTION
Use under unallowable environmental conditions
Risk of injury due to damage to the product
Do not expose the product to unallowable environmental conditions.
►
If the product has been exposed to unallowable environmental conditions, check it for dam
►
age.
If damage is apparent or in case of doubt, do not continue using the product.
►