The fault is shown by an error message in the display (C) as following if the error occurs, e.g. Error 4.
Proceed as follows in such cases:
 Switch the device switch (A) off.
 Remove the stirrer tool and remove the instrument from the assembly.
 Reduce the speed and switch on (instrument switch (A)) the device without the stirrer tool.
Error
Cause
Error 2
Motor current sensor fault
Error 3
Temperature inside instrument is too high
Error 4
Motor blockage or overload
Error 8
Speed sensor fault or overload
Error 21
Safety relay fault
If the actions described fails to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the service department
- Send the instrument for repair, including a short description of the fault.
In accordance with IKA
®
warranty conditions, the warranty period
is 24 months. For claims under the warranty please contact your
local dealer. You may also send the machine direct to our factory,
enclosing the delivery invoice and giving reasons for the claim. You
will be liable for freight costs.
R 2722
H-stand
R 2723
Telescopic stand
R 270
Cross sleeve
R 271
Cross sleeve
RH 5
Strap clamp
FK 1
Flexible coupling
R 301
Stirrer shaft protection
R 301.1
Support holder
R 1342
Propeller stirrer
R 1345
Propeller stirrer
R 1381
Propeller stirrer
R 1382
Propeller stirrer
R 1385
Propeller stirrer
R 1388
Propeller stirrer
R 1389
Propeller stirrer, PTFE
R 1311
Turbine stirrer

Error codes

Effect
motor off
motor off
motor off
motor off
motor off

Warranty

The warranty does not cover worn out parts, nor does it apply
to faults resulting from improper use, insufficient care or mainte-
nance not carried out in accordance with the instructions in this
operating manual.

Accessories

PC 1.1
H 70
H 62.51
H 66.51
RB 1
OS 1.0
USB cable micro A – micro B 2.0
USB cable A – micro B 2.0
Permitted IKA
stirrer tools
®
max. speed
(rpm)
≤ 2000
R 1312
≤ 800
R 1313
≤ 2000
R 1375
≤ 2000
R 1376
≤ 800
R 1330
≤ 400
R 1331
≤ 800
R 1333
≤ 2000
Solution
- Switch off the instrument
- Switch off the instrument and allow it to cool down
- Switch off the instrument
- Decrease the load on the motor and restart again
- Switch off the instrument
- Switch off the instrument
Cable
Extension cable
Stainless steel sensor
Stainless steel sensor, glass-coated
Battery pack
Power supply
Turbine stirrer
Turbine stirrer
Paddle stirrer
Paddle stirrer
Anchor stirrer
Anchor stirrer
Anchor stirrer
max. speed
(rpm)
≤ 2000
≤ 800
≤ 800
≤ 800
≤ 1000
≤ 1000
≤ 800