4.3 MODES

STANDBY MODE
The Penguin Nutritional Warmer will be in Standby mode when it is turned on
without well in use.
WARMING MODES
Room Temperature/Refrigerator to Feed
Feed Temp: range of 90-100°F/32-38°C
Average Time: 5-12 minutes depending on volume
Frozen to Feed
Feed Temp: range of 90-100°F/32-38°C
Average Time: 10-18 minutes depending on volume
Thaw
After Thaw Temp: range of 38-45°F/3-7°C
Time: 15-25 minutes depending on volume
READY TO FEED MODE
The feeding is ready when:
Ready LED is flashing green
The warmer emits a ready tone.
(Ready tone will periodically emit during Ready to Feed mode).
Onscreen display reads:
Ready to Feed AT x MIN
REMAINING HOLD TIME x MIN
The Remaining Hold Time displays a countdown of the remaining time. The warmer
will keep the feeding in a safe temperature range for 30 minutes (until turned off,
returned to Standby mode, or extends to Overtime mode).
OVERTIME MODE
After 30 minutes in Ready to Feed, the warmer will enter Overtime mode. The
Overtime screen will remain on until the unit is turned off or is returned to Standby
by pushing the Warm/Off button.
Note: As the manufacturer, we recommend not using the feeding after
the 30 minute hold time, but that feeding should be handled according
to the hospital policy.