Electromagnetic
14
compatibility
Manufacturer's guide and declaration – electromagnetic immunity
The "CCD mono video camera" equipment is intended for use in the electromagnetic environment
specifi ed below.
The user will have to make sure that it is used in this environment.
Immunity test
Electrostatic
discharges
IEC 61000-4-2
Transient voltage
in bursts
IEC 61000-4-4
Voltage surges
IEC 61000-4-5
Dips, short
interruptions
and variations in
thepower voltage
IEC 61000-4-11
Magnetic fi eld at
the
network's
frequency
(50/60 Hz)
Note:
UT is the rated value of the power supply voltage applied during the test.
24
IEC 60601
Compliance
Level of gravity
level
± 6 kV in Contact
± 6 kV
± 8 kV in air
± 8 kV
± 2 kV for power
± 2 kV
lines
± 1 kV for input/
± 1 kV
output lines
Di erential mode
± 1 kV
± 1 kV
Common mode
± 2 kV
± 2 kV
• < 5 % UT –
< 5 % UT
for 10 ms
10 ms
• 40 % UT –
< 40 % UT
for 100 ms
100 ms
• 70 % UT –
< 70 % UT
for 500 ms
500 ms
• < 5 % UT –
< 5 % UT
for 5 s
5 s
3 A/m
± 6 kV via
contact
± 8 kV via air
Electromagnetic environment - guide
The fl oor must be made from wood,
concrete or tiling. If the fl oor is covered
with a synthetic material, the relative
humidity must be less than 30%.
The quality of the main power supply
must be that of a typical hospital or
commercial environment.
The quality of the main power supply
must be that of a typical hospital or
commercial environment.
The quality of the main power supply
must be that of a typical hospital or
commercial environment. If the user of
the "CCD mono video camera" reference
equipment requires it to be able to
continue to function during interruptions
in the main power supply, it is
recommended that the "CCD mono video
camera" reference equipment is supplied
by an uninterruptible power supply or a
battery.
The magnetic fi eld at the network's
frequency must be at a level characteristic
of a location in a typical hospital or
commercial environment.