AIR/WATER COOLED HOUSING
The air/water cooled housing shown below allows the sensor to withstand high ambient
temperatures. It is equipped with two 1/8" BSP fittings. Water temperature should be 10°C to 27°C
for efficient cooling. Chilled water below 10°C is not recommended. To avoid condensation, the
air purge collar should be used with the water-cooled housing. Water flow rate should be no more
than 0.5 to 1.5 litres/min.
1/8 BSP water/air connections
19
AIR PURGE COLLAR
The air purge collar below is used to keep dust, fumes, moisture, and other contaminants away
from the lens. It must be screwed in fully. Air flows into the 1/8" BSP fitting and out of the front
aperture. Air flow should be no more than 5 to 15 litres/min.
Clean or 'instrument' air is recommended.