INSPECTION OF PATIENT HOISTS
IDENTIFICATION OF THE MEDICAL EQUIPMENT
CATEGORY:
BRAND / MODEL / TYPE:
SERIAL NUMBER:
DATE OF MANUFACTURE:
VISUAL INSPECTION (According to NF EN ISO 10535)
Availability and good condition of the instruction manual, loading plate, identificación plate / labe-
lling / pinting.
Good general condition (structure, base, articulations, welds, attachments, etc ) Absence of rus-
ting.
Good general cleanliness.
Good stability of the hoist and good operation of the base opening / closing.
Checking of the tightness of screws (suspension pin of the lifting arm, weighhing module attache-
ments, etc..) absence of space.
Good condition of slings and cladle or hammock, including the attachment system.
Good rotation of the gripping bar and good condition of the hooks.
Good operation of the wheels and brakes (swivelling, running, locking).
Good lubrication of articulations (absence of noise, nuisance, squeaking and hard points).
Good condition of electric cables, control boxes, connectors and actuators (jacks, motors ).
Good condition of controls and indicator lights.
SAFETY INSPECTION (According to NF EN ISO 10535)
Looking out of operational functions and correct operation of visual and audible alarms.
Emegency stop operates correctly.
Battery and charger opérate correctly.
Obstacle stop operates correctly.
Correct operation of the emergency descent.
Test at máximum load (1 complete cycle).
INSPECTION OF THE AMPLITUDE OF MOVEMENTS
Satisfactory operation of the actuator (elevation/descent)
Maximum and mínimum height (according to the manufacturer specifications)
ELECTRICAL SAFETY (According to EN 60601-1)
Tester Model:
Continuity:
Breakdown:
Name:
Date:
Serial number:
INSPECTOR
ESTABLISHMENT:
SIGNATURE:
ESTABLISHMENT
DEPARTMENT/LOCATION:
NA: NOT APPLICABLE C: CONFORM
NA
NA
NA
NA
.
NC: NON CONFORM
C
NC
C
NC
C
NC
C
NC