Mini Laboratory pumps N 86.18, N 811.18
1
Outlet valve
2
Inlet valve
3
Transfer chamber
4
Diaphragm
5
Eccentric
6
Connecting rod
7
Pump drive
Translation of original Operating Instructions, english, KNF 121435-121438 11/12
Function Diaphragm Pump
Fig. 4: Pump head
Diaphragm pumps transfer, compress (depending on pump ver-
sion) and evacuate gases and vapors.
The elastic diaphragm (4) is moved up and down by the eccentric
(5) and the connecting rod (6). In the downward stroke it aspirates
the gas to be transferred via the inlet valve (2). In the upward
stroke, the diaphragm presses the medium out of the pump head
via the outlet valve (1). The transfer chamber (3) is hermetically
separated from the pump drive (7) by the diaphragm.
Design and Function
13
DE