Connection of interfaces
Observe the relevant connections
NOTE
(see Fig. 1).
O: Hose connection d= 8 mm OUTLET
Link this connection to the emission condenser with a vacuum
hose, or fit a sound absorber to the end.
Put the end of the hose in the
NOTE
extractor hood!
Check there is a free outlet on the
pressurised side!
Do not use a throttle on the
pressurised side and do not close
the outlet! Connect the exhaust
line to this connection.
K: Connection for vacuum controller VC 10 (Mini DIN)
You can connect the vacuum pump and the vacuum controller
VC 10 or the rotary evaporator RV10 auto with the analogue
connection cable MVP 10.100 for precise speed-controlled
vacuum control.
The vacuum controller detects the pump and switches to
speed-vacuum control mode. Two-position control is deactivated.

Configuration

Observe the operating instructions
NOTE
for the devices.
Two-position control:
IKA Vacstar digital in standalone operation
Vacuum created without a controller through simple suction, no
vacuum target value control possible.
The pump speed can be changed by manually setting the speed
using the rotating/pressing knob (C) of the vaccum pump.Manual
1 Recipient (load, e.g. rotary
evaporator, reactor)
2 Vacuum separator (e.g. Woulff
bottle)
5 Vacuum valve/ball valve
7 Emission condenser
10
L: USB interface
Connect the vacuum pump IKA Vacstar digital to a PC with a USB
A - USB B cable. Any device software updates can be loaded using
the IKA FUT software tool.
M: RS 232 interface
You can connect the IKA Vacstar digital vacuum pump to a PC
using an RS 232 interface cable. The pump can be operated in
conjunction with other devices with labworldsoft
device software. For further information see the chapter "Interfaces
and outputs".
N: Water valve connection
Connect the optional water throttle valve RV 10.5001 to the
diaphragm vacuum pump. The water flow to the emission
condensor is controlled using the water valve. The valve is opened
as soon as the pump is switched on.
I: Hose connection for suction line d= 8 mm INLET
Connect the suction line to this connection.
Link this connection to the recipient (rotary evaporator condenser,
laboratory reacor etc.) with a vacuum hose.
P: Power supply cable connection
Check that the voltage information on the rating plate matches
your mains supply. Connect the power supply cable to supply
electricity.
activation of the vacuum valve or ball valve.
Rough vacuum control is possible, for example, with the vacuum
control valves VCV 1 and VCV 2, by mixing in air from outside.