4.3.
DESCRIPTION OF LCD DISPLAY
Item
Symbol
Data HOLD function activated
1
HOLD
2
The symbol will light on when there is a voltage over 50VAC/120VDC
3
Clockwise phase sequence symbol
Anticlockwise phase sequence symbol
4
Low battery indication
5
LCD display
6
Voltage and Resistance units
V, k
DC Voltage measurement
DC
7
AC Voltage measurement
AC
Negative polarity on DC voltage measurement
4.4.
INITIAL AUTOTEST
Before starting any measurement perform the following checks:
Use the instrument on a known voltage source
The "
" LED switches on when a voltage above 50VAC/120VDC is applied, even if
the battery is low or has been removed
Before and after the use of instrument check as follows:
Short the L1 and L2 probes for approx. 4-6s. The below actions must be observed
All the LEDs except the low impedance test switch on
The audible signal will sound
All segments of LCD display and backlight switch on
Fig. 3: Description of LCD display
EN - 7
Description
HT10