Model
Typology (MDD 93/42/EEC)
Classification UNI EN ISO 10079-1
Main Voltage
Power consuption
Fuse
Maximum suction aspiration (without jar)
Maximum flow (without jar)
Weight
Dimension
Functioning (to 35°C and 110 % operating voltage)
Working condition
Conservation condition
Please note tecnica specifications may vary upon the manufactures's discretion!
The surgical aspirator NEW ASKIR 20 is intended for use in the electromagnetic environment specified below.
The customer or the user of the surgical aspirator NEW ASKIR 20 should assure that it's used in such an environment.
Emissions test
Irradiated / Conducted emissions
CISPR11
Irradiated / Conducted emissions CISPR11
Harmonic emissions IEC/EN 61000-3-2
Voltage fluctuations / flicker emissions
IEC/EN 61000-3-3
Any re production of this instructions for use is forbidden without CA-MI Srl approval
TECHNICAL CHARACTERISTICS
~
Hz
I
0
Guidance and manufacturer's declaration – electromagnetic Emissions
Compliance
Complies
Regolable from -75kPa (-0.75 Bar) to -10kPa (-0.10 Bar)
Room temperature:
Room humidity percentage:
Altitude:
Room temperature:
Room humidity percentage:
SYMBOLS
Class 2 isolation equipment
CE marking in conformity with EC directive 93/42/CEE and subsequent changes
Manufacturer: CA-MI Srl - Via Ugo La Malfa n°31, 43010 Pilastro (PR) Italy
Warning, consult the instruction manual
Type B equipment
Fuse
To preserve in place coolness and dry land
Conservation temperature: - 40 ÷ 70° C
Alternate current
Mains frequency
ON
OFF
DEHP Phthalates (Suction catheter)
Group 1
The surgical aspirator NEW ASKIR 20 only used RF energy only for its
internal functioning. Therefore, its RF emissions are very low and are
not cause interference in proximity of any Electronic appliances.
Class [B]
The surgical aspirator NEW ASKIR 20 can be used in all environments,
including domestic and those connected directly to the public mains
Class [A]
distribution that supplies power to environments used for domestic
scopes.
11
NEW ASKIR 20
Medical Device Class II a
HIGH VACUUM / LOW FLOW
230 V ~ / 50 Hz
184 VA
F 1 x 1.6A 250 V
-75kPa (- 0.75 bar)
16 l/min
2.5 Kg
350 x 210 x 180 mm
NON-STOP OPERATED
5 ÷ 35 °C
30 ÷ 75 % RH
0 ÷ 2000m s.l.m
- 40 ÷70 °C
10 ÷ 100% RH