3.
The Dock's display will turn off for a few seconds while the unit reboots and reprograms its embedded
firmware. Once the display turns on, verify that the new firmware revision is correctly reported. (The
image below is for illustration purposes only. Your firmware P/N and version may differ):
4.
The display will prompt you to remove the Programming Cassette:
Once the Programming Cassette is removed, the Accula™ Dock will restart with the new
firmware installed.
D. Maintenance
Cleaning the Accula™ Dock
Regular cleaning of the Accula™ Dock is recommended. The Dock should be cleaned every day that it is
used, and prior to disposal or shipment or servicing.
A used Dock may be biologically contaminated. Prior to disposal, or shipping for service, the
F
Dock must be thoroughly cleaned.
To reduce the risk of damage to the unit, avoid use of sprays in the vicinity of the Dock.
Y
Prior to cleaning the Accula
Y
damage to the unit.
Mesa Biotech recommends cleaning the Dock each day it is used with CaviWipes™ germicidal disposable
wipes, (EPA Registration No: 46781-8). The cleaning procedure below has been validated using these wipes.
CaviWipes™ germicidal disposable wipes contain the following active ingredients:
Isopropanol: 17.2% and
Diisobutyl-phenoxy-ethoxyethyl dimethyl benzyl ammonium
chloride: 0.28%.
For information on purchasing CaviWipes™ germicidal wipes, contact your authorized distributor.
If Cavi Wipes are unavailable, other cleaning solutions may be used such as 10% bleach, or 70% ethanol.
Procedure:
1.
Unplug the power cord from the Dock.
2.
Close the Dock lid.
3.
Using a CaviWipes™ germicidal disposable wipe, gently wipe the outer surfaces of the Dock, removing
any gross debris. Do not press the wipe against the open vents of the back or sides of the Dock.
4.
Using a new CaviWipe, wipe the front of the Dock twice top to bottom, then twice left to right. Follow
this procedure for the back, top and bottom of the Dock.