INSTRUCTION MANUAL
J
.P. SELECTA s.a.u.
CODE 80300 REV A
9.1 Reset for overtemperature
The equipment has a separate safety device to prevent that the heating element tem-
perature rises uncontrolled, for example, when it runs out of water.
In that case, the mechanism disconnects the heating element power and, once cooled,
it must be reset in order to continue using the computer.
Press the reset button on the back side of the computer.
9.2 Setting the «Offset» temperature reading
Only use this adjustment if a deviation is detected between the temperature indicated
by the thermostat and the one indicated by a pattern thermometer.
Simultaneously press the keys
The symbol "
For example: if the equipment indicates 40.0ºC and the pattern indicates 40.5ºC, adjust
the value: -0.5ºC
10 WARRANTY
This product is guaranteed for one year. It does not cover damages caused by an incorrect
use or causes beyond the control of ICC.
Any manipulation of the equipment by personnel not authorized by ICC automatically
cancels the guarantee.
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
Sep2016
(It can be modifi ed without notice)
o
" will appear in the lower display and adjust the value with
to set the "Offset" value for 2 seconds-
Tel 34 937 700 877 Fax 34 937 702 362
Page: 8
.