Measuring
earth/ground-connection resistance „R
" and the tripping time „t
" (not for
A
A
measurement without tripping) and then again „U
" into the display. For a new
L
R
measurement press key
„Start".
To return to the original measuring mode RCD for displaying the mains voltage
U
, briefly turn the central selector.
LPE
ATTENTION! To be certain that the fault current-protection circuit is operative, the
measurement must be effected at the first measuring point (outlet, device ...) of
each FI current circuit with tripping of the FI-Circuit breaker! Measuring without
tripping is only permitted for all measuring points connected in parallel.
28