Ditel JR-P Manual Tecnico página 63

Indicador universal para entrada proceso y temperatura
Ocultar thumbs Ver también para JR-P:
Idiomas disponibles

Idiomas disponibles

AVAILABLE KEYBOARD FUNCTIONS
In addition to already known functions used to browse through the configuration menus and submenus, introduce
and/or modify existing values and parameters, the instrument provides some more added functions.
MAX/MIN and RESET functions
This device detects and stores in memory maximum and minimum values reached by the input signal. This values
are kept in memory although power supply is desconnected. When pressing repeatedly SHIFT key, MAX/MIN
function shows saved maximum and minimum values in display since last RESET function activation.
In order to differentiate this values indication from a mode RUN indication, decimal point blinks during the time
these values are showed. The unit automatically switches back to RUN mode after 15 seconds have elapsed since
the last key press.
First SHIFT key pressing shows "MAH" in display followed by the maximum value, a second pressing now shows
"Min" followed by the minimum value and finally, a third pressing shows "run" to back again in an instant to RUN
mode.
RESET function activates when visualizing maximum or minimum values SHIFT key is pressed for at least 5
seconds. If maximum is the displayed value, current input signal value will replace the previous maximum saved
value. In the same way, current input signal will replace saved minimum value while is the minimum the displayed
value.
Direct access to setpoints value
Pro
SEt1
00.00
SEt2
00.00
Return to default configuration
Pro
5s
00
74
SERIE KOSMOS
If 2RE output option is installed, it is possible to access to setpoints value
configuration without having to enter main menu.
To access this submenu, from RUN mode and after ENTER key is pressed, simply
press UP key while "Pro" is displayed.
FIRST SETPOINT VALUE:
SEt1:
Setpoint 1 value indication.
00.00:
Value entering in counts within available model display range.
SECOND SETPOINT VALUE:
SEt2:
Setpoint 2 value indication.
00.00:
Value entering in counts within available model display range.
To access this menu from RUN mode, press ENTER key and while display shows
"Pro" press again ENTER for at least 5 seconds.
Display shows now "00" and '74' code must be introduced through SHIFT and UP
keys.
Finally press ENTER to validate configuration and back to RUN mode.
www.ditel.es
JR-P / JR20-P
63
loading

Este manual también es adecuado para:

Jr20-pJunior-pJunior20-p