Troubleshooting
Failure
Failure 4:
Handheld sprayer operates normally but no
water or air is sprayed out.
Failure 5:
Program chair fails to store data
Failure 6:
There is water or oily dirt in air hose.
Mantenimiento
52 Maintenance
Troubleshooting
Inspections:
1. First check whether the sprayer operates currently in dry grinding mode
where no water or air is sprayed by operating the change-over switch on it.
For more details, see the operating instructions.
2. Check the water/air hose line is in good connection in instrument panel. If the
hose line is in good condition, then check whether there is DC 24 V output at
water/air circuit solenoid valves;
if yes, the solenoid valve may be broken, and replace solenoid valve; if not,
measure voltage output at reserved port J33, if there is a voltage output,
replace treatm ent platform control board, if not, inspect the power supply in
side case.
3. Foot-operated micro switch:
Check whether the foot-operated micro switch resets norm ally;
if not, replace the micro switch.
4. Replace side case control board.
Inspections:
Generally, failure of program chair storage function is resulted from dam age to
angle sensor, replace the corresponding angle sensor.
Solution:
Check air supply source and air filter,
clean or replace relevant components.