Installing The Reliant Scale - Invacare Reliant 450 RHL450-1 Manual Del Usuario

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

11 ACCESSORIES

Installing the Reliant Scale

For this procedure, refer to FIGURE 11.2.
WARNING
Patient and sling MUST be removed from the lift during ALL installation procedures.
1.
Position the load cell assembly of the Reliant Scale into the boom mounting bracket. Refer to Detail "A" in FIGURE 11.2.
Use 1/4-inch nylon washer.
2.
Secure the Reliant Scale to the boom mounting bracket with a shoulder bolt, two nylon washers and a locknut. Securely tighten. Refer to Detail
"A" in FIGURE 11.2 for washer orientation.
Ensure the shaft of the shoulder bolt passes through both sides of the boom mounting bracket.
3.
Insert hanger bar pin with two nylon washers through the hanger bar. Refer to Detail "B" in FIGURE 11.2.
4.
Align the mounting holes in the hanger bar pin with the mounting holes in the load cell assembly. Refer to Detail "B" in FIGURE 11.2.
5.
Secure hanger bar pin to the load cell assembly with the provided mounting screw and locknut. Securely tighten. Refer to Detail "B" in
FIGURE 11.2.
WARNING
After ANY adjustments, repair or service and BEFORE use, make sure all attaching hardware is tightened securely - otherwise injury or
damage may occur.
6.
Weigh the patient. Refer to Operating the Scale on page 54 and Weighing the Patient on page 55.
Manual/Electric Portable Patient Lift
52
Part No. 1078987
Tabla de contenido
loading

Tabla de contenido