WIKA BCS10 Manual De Instrucciones página 15

Ocultar thumbs Ver también para BCS10:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

6. Commissioning, operation
Select an appropriate adapter and screw it
together with the test item.
Connect the test item via the quick coupling to
the metal flexible hose.
EN
For instruments with calibration valve, also
observe the mounting instruction of the
respective operating instructions.
Connect the other side of the metal flexible hose
via the quick coupling to the calibration system.
Check that the switching for pressure or vacuum
generation (arrow) is set to pressure.
ATTENTION!
The construction of the calibration system is not
designed for vacuum generation. Always leave
the switching valve set to pressure generation.
Never adjust the switching valve when the
system is under pressure. Switch only when the
drain valve is open.
Close the pressure relief valve.
Turn the pressure relief valve clockwise until the
valve is closed. No hard stop can be felt.
WIKA operating instructions calibration system, model BCS10
15
Tabla de contenido
loading

Tabla de contenido