Operating Conditions - WIKA ACS-10 Manual De Instrucciones

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

9. Specifications
Components
Maximum connection pressure
CON2 (filling and emptying)

Operating conditions

Medium temperature
Ambient temperature
Storage temperature
Relative humidity
Ambient pressure
Vibration resistance
Shock resistance
Ingress protection
Closed case, transport
Open case, operation
Transport
Maintenance
For further specifications see WIKA data sheet SP 60.15 and the order documentation.
WIKA Betriebsanleitung Kalibriersystem, Typ ACS-10
10 bar abs.
5 ... 40 °C [41 ... 104 °F]
5 ... 40 °C [41 ... 104 °F]
-10 ... +60 °C [14 ... 140 °F]
10 ... 90 %
600 ... 1,060 hPa
EN 30786-2, annex A.2.7
EN 60068-2-31, chapter 5.1.3.3 (overturning) and chapter
5.2 (free fall)
IP65
IP40
The instrument may only be transported in Transport mode
(pressure in the instrument < 2 bar abs.).
Recommended recalibration of the reference sensor in a
WIKA service hub after one year or 7,000 measurements