5. Commissioning, operation
5.9 Recalibration of a gas density indicator
Model ACS-10 can also recalibrate gas density indicators. For this purpose, specific
pressures are approached, which the operator must then read visually on the gas
density indicator and confirm in the model ACS-10. Before doing so, a new instrument
must be created in the data management, as is the case with other types of test items
(→ see chapter 5.4 "Creating a new measuring location").
Then save the instrument setup and press the "Test Device" button. After entering the
height difference, the measurement starts.
EN
Model ACS-10 approaches the calibration point of the instrument and then asks for the
pointer position of the test item.
82
WIKA Betriebsanleitung Kalibriersystem, Typ ACS-10