8. Faults / 9. Return and disposal

8. Faults

GB
Indication Type of fault
OL
-OL
No display or
undefinable
characters, the
instrument does
not respond to
key press

9. Return and disposal

9.1 Return

22
Battery is low, functioning is only
guaranteed for a short period of time
Reading is significantly above or
below the measuring range
Battery is dead.
System error
The CEP1000 is defective
CAUTION!
If faults cannot be eliminated by means of the measures listed above,
the instrument must be shut down immediately, and it must be ensured
that pressure and/or signal are no longer present, and it must be
prevented from being inadvertently put back into service.
In this case, contact the manufacturer.
If a return is needed, please follow the instructions given in chapter
"9.1 Return".
WARNING!
Residual media in dismounted loop calibrator can result in a risk to
persons, the environment and equipment. Take sufficient precautionary
measures.
WARNING!
Strictly observe when shipping the instrument:
All instruments delivered to WIKA must be free from any kind of hazard-
ous substances (acids, bases, solutions, etc.).
Measures
Insert new batteries
Test: if the measurement parameter
is within the approved measuring
range of the instrument?
Insert new batteries
Turn off the CEP1000, wait for a
short period of time, switch on again
Send in for repair
WIKA operating instructions, model CEP1000