► Rinsing the filter cartridge
⇒ Run the flushing hose into a suitable container (e.g., a bucket) or into a drain.
en
⇒ Open the purge valve (5) (OPEN, Fig. 3).
⇒ Turn on the water supply.
Water escapes from the flushing hose as soon as the filter cartridge has been inserted!
⇒ Insert the filter cartridge into the filter head.
Insert the filter cartridge with its marking at insertion mark A (Fig. 5) and screw it in as far as it will
go, operating mark B (Fig. 5).
⇒ Vent and rinse the filter cartridge.
- Filter cartridge M: Flush with min. 5 liters of water
- Filter cartridge L:
⇒ Close the purge valve (5) (CLOSE, Fig. 4).
⇒ After the initial installation of the filter system, vent and flush the water drain hose and the POWER
steamer 2 with at least 2 liters of water. To do this, open the service opening on the POWER steamer 2
and run the rinsing program 3 times (see operating instructions for the POWER steamer 2).
After installing the system and inserting or replacing a filter cartridge, check all components for
leak-tightness. Water must not leak out at any location.
4.2
Installing the programming and display unit
To operate the programming and display unit, you will need 2 batteries (type AA). These are not
included in the scope of delivery!
The programming and display unit can either be attached to the filter head (2)
of the filter unit using the holder (11) or to a wall using the wall bracket provid-
ed (12).
⇒ Push the holder (11) onto the filter head (2) until it snaps into place.
or
⇒ Screw the wall bracket securely to a wall using suitable flat-head screws,
max. shank diameter 4 mm, or with the adhesive pads on the back.
⇒ Open the battery compartment and insert batteries, 2x type AA.
⇒ Pay attention to correct polarity.
⇒ Hang the programming and display unit in the holder.
⇒ Insert the connection cable of the sensor unit into the port (20).
In the case of assembly using the wall bracket, the connection cable can be
pressed into the cable guide (21) to prevent the connection plug from being
unintentionally pulled out.
4.2.1
Entering the filter capacity / resetting the counter
⇒ Determine the capacity of the filter cartridge used for the local carbonate hardness (see table in the
installation section).
⇒ Enter the capacity:
Key
Flush with min. 10 liters of water
Press the "MENU" key until CAP SET is shown.
- Display: Set filter capacity.
Press and hold the "MENU" key for approx. 5 seconds.
- The active digit flashes.
Press the "MODIFY" key:
- Set the active digit.
Press the "MENU" key:
- Confirm the active digit and move to the next digit.
- The active digit flashes (repeat the procedure until the last digit).