Specifiche tecniche e caratteristiche prestazionali del dispositivo
Dead Space
nonrebreathing valve
adult mask
child mask
infant mask
Adult model
Child model
Infant model
Stroke volume of 1350 mL can be achieved using two hands
Minimum Cycle Rate
Adult - 20 breaths/min
Child - 20 breaths/min
Infant - 40 breaths/min
The performance characteristics for Manual Resuscitators will vary from user to user depending on a variety
of factors: ambient temperature, patient lung compliance, ventilatory frequency, size of operator's hands.
Maximum Ventilatory Rates
Adult
Child A
Child B
Infant A
Infant B
The results were obtained under the following conditions:
Adult:
VT-600mL, Compliance 0.02L/cmH
Child A:
VT-70mL, Compliance 0.01L/cmH
Child B: :VT-300mL, Compliance 0.01L/cmH
Infant A: :VT-20mL, Compliance 0.001L/cmH
Infant B: :VT-70mL, Compliance 0.01L/cmH
The correct ventilation frequency may vary, please follow current ventilation frequencies as recommended
by AHA.
Range of delivery pressure
Adult: 60±10 cmH
O maximum (can be overridden by operator)
2
Child and Infant: 40±5 cmH
ENGLISH
7 mL
150 mL
95 mL
28 mL
Bag volume
Stroke volume
1600 mL
700 mL
500 mL
300 mL
280 mL
150 mL
-18°C(0°F)
Cycle Rate
20
30
20
60
40
2
O, Resistance 20 cmH
2
O, Resistance 20 cmH
2
O maximum (can be overridden by operator}
2
Pressure Relief Valve
Child and infant 40±5 cmH
Adult 60±10 cmH
Reservoir volume
2500 mL
2500 mL
500 mL
Oxygen Concentration
with reservoir
without reservoir
Temperature
22°C(72°F)
Cycle Rate
20
30
20
60
40
O, Resistance 20 cmH
O/L/s
2
O, Resistance 20 cmH
2
O, Resistance 400 cmH
2
2
O
2
O
2
Suitable body weight
>30 kg
7-30 kg
<7 kg
99%
45% (adult and child models)
90% (infant model)
50°C(122°F)
Cycle Rate
20
30
20
60
40
O/L/s
2
O/L/s
2
O/L/s
2
O/L/s
24