Problem
Pump has no
The pump rod 5 is not seated
suction or too
properly in the pump
little suction
diaphragm 7
Pump diaphragm 7 missing,
fitted incorrectly or damaged
Lip valve 9 is incorrectly
positioned
Lip valve 9 missing, fitted
incorrectly or damaged
Breast milk container 11 is full
and lip valve 9 is touching the
expressed milk
No vacuum is produced
Pump is being held at an angle Hold the pump as upright as possible.
Medical product
Temperature range
- Operation
- Storage
If at some stage the appliance is no longer
fit for use, the parts can be disposed of
along with normal household waste.
In
addition
to
obligations arising from the contract to
purchase, we as the manufacturers provide
a warranty of 24 months from purchase of
the appliance as long as the appliance is
used properly and the operating instructions
11. Help with problems
Reason
12. Technical data
13. Disposal
14. Warranty
the
seller's
warranty
Make sure that the pump rod 5 is fully inserted in the
hole in the pump diaphragm 7 (see "7. Reassembling
the pump (Figures I to P)" on page 18).
Insert pump diaphragm 7 correctly (see
"7. Reassembling the pump (Figures I to P)" on
page 18). If anything is damaged, please contact a
NUK Service Center (see "Contact addresses" on
page 90).
Position lip valve 9 correctly (see "7. Reassembling
the pump (Figures I to P)" on page 18).
Check that the lip valve 9 is correctly positioned.
If anything is damaged, please contact the NUK
Service Center (see "Contact addresses" on
page 90).
Change the breast milk container 11.
Make sure that the breast shield 2 forms a
complete seal around the breast.
Make sure that the silicone cushion 1 is inserted
correctly and is pulled over the edges of the
breast shield.
Regulation (EU) 2017/745
+6 °C to +40 °C
0 °C to +40 °C
are followed. Evidence of the date of
purchase and the type of appliance should
be provided by means of a receipt.
The
purchaser's
unaffected by this manufacturer warranty.
22
Remedy
statutory
rights
are