MANUAL DE INSTRUCCIONES CODIGO 80201 REV A
ISO
Isopotential p oint selection
s cr ee n
MTC
Entering measuring
temperature screen.
25.0ºC
MTC
Enter measuring
te mpe ra tur e.
ASD
Automatic p H-meter
disconection.
J
.P. SELECTA s.a.
6.10 Isopotencial point selection
Select suitable isopotential point to measure pH value of pure water,
super pure water and boiler water.
Usually, isopotential point is set to 7.00pH to measure general water
solution.
To measure pure or supper pure water set 12.00pH
For ammonia water set 17.00pH.
6.11 Measuring temperature. Automatic compensation.
The pH-2004 includes ATC (Automatic temperature correction). This
means that the reading is corrected if the measuring temperature is
different from 25ºC.
The measuring temperature is the one the temperature probe reads or if
the temperature probe is NOT connected it is the temperature set by
the keypad .
Setting the measuring temperature by the keypad:
(Temperature probe NOT connected on the rear p anel socket)
•
Press «MODE/MEAS» and select «MTC» using arrow keys.
•
Validate with «ENT ER».
•
Use keys «
•
Validate with «ENTER» and press «MODE/MEAS» to return.
Note: If the measuring temperature set have a value far from the actual
sample temperature the pH reading could have considerable error.
6.12 Automatic switching Off
Sets the automatic disconnection after 2h of no operation.
•
Press «ENTER».
•
Use keys «
ON :
OFF:
6.13 Buffer solutions.
For a certified calibration, some standard buffers must be used, As the
ones described on 1.2.
Clean, rinse, with water, the electrode before to immerse into buffers.
• Buffer solutions are irritant chemicals. Avoid any contact with eyes and
use a lab coat to protect your wear.
• In case of chemicals be in contact with skin, clean, quickly, with abundant
w a te r.
• To prepare the buffer solutions just take a 300-400 ml flask, and solve
the pounder bag into 250ml of distilled water. Stirring until solved.
• Keep this solutions in a closed flask. Do not use this buffer for calibration
after 1 moth after preparation.
6.14 RS-232 Communication
Use the communication software to capture reading on the computer.
Connect the communication wire on a free RS-232 COM port on your
comp uter. (Sp eed 9600)
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
FEB 2013
(Sujetas a modificaciones sin previo aviso)
» «
» to select the temperature.
»: y «
»
Automatic switch Off
No automatic switch Off
Atten tion
Tel 34 937 700 877 Fax 34 937 702 362
Pág..: 18