Cleaning And Storing The Device And Cuff; Technical Details - Beurer BC 44 Manual De Instrucciones

Ocultar thumbs Ver también para BC 44:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 39
• the pump pressure is higher than 300 mmHg (
pears on the display)
• pumping up takes longer than 160 seconds (
on the display)
• there is a system or device error (
appears on the display)
• the batteries are almost empty
In such cases, repeat the measurement. Ensure that you do
not move or speak during the measurement. If necessary,
reinsert or replace the batteries.
Technical alarm – description
Should the recorded blood pressure (systolic or diastolic)
lie outside the limits specified in the section "Technical
specifications", the technical alarm will appear on the dis-
play indicating either "Hi" or "Lo". In such cases, you should
seek medical assistance and check the accuracy of your
procedure.
The limit values for the technical alarm are factory set and
cannot be adjusted or deactivated. These alarm limit val-
ues are accorded second priority under the standard IEC
60601-1-8.
The technical alarm is a non-locking alarm and must not be
reset. The signal shown on the display will disappear auto-
matically after about 8 seconds.
ap-
appears
,
,
or
.
23

9. Cleaning and storing the device and cuff

• Clean the device and cuff carefully using a slightly damp
cloth only.
• Do not use any cleaning agents or solvents.
• Under no circumstances hold the device and cuff under
water, as this can cause liquid to enter and damage the
device and cuff.
• If you store the device and cuff, do not place heavy ob-
jects on the device and cuff. Remove the batteries. The
cuff line should not be bent sharply.

10. Technical details

Model no.
BC 44
Measurement
Oscillometric, non-invasive blood pres-
method
sure measurement on the wrist
Measurement
Cuff pressure 0 – 300 mmHg,
range
systolic 60 – 260 mmHg,
diastolic 40 –199 mmHg,
Pulse 40 –180 beats/minute
Display accuracy Systolic ± 3 mmHg, diastolic ± 3 mmHg,
pulse ± 5 % of the value shown
Measurement
Max. permissible standard deviation
inaccuracy
according to clinical testing:
systolic 8 mmHg /diastolic 8 mmHg
Dimensions
L 87,8 mm x W 65,8 mm x H 28,6 mm
Tabla de contenido
loading

Tabla de contenido