P P r r e e v v e e n n t t i i v v e e m m a a i i n n t t e e n n a a n n c c e e
Remove product from service before you perform preventive maintenance inspection. Check all items listed during annual
preventive maintenance for all Stryker Medical products. You may need to perform preventive maintenance checks more
frequently based on your level of product usage. Service only by qualified personnel.
N N o o t t e e
• Clean and disinfect the exterior of the support surface before inspection, if applicable.
• If excessive wear is observed on the cover or the fire barrier, the recommendation is to replace the cover.
Inspect the following items:
All fasteners are secure
Zipper, snaps, and covers (top, bottom, and fire barrier) are free of tears, cuts, holes, or any other damage
Support surface cover labels are legible, adhere, and free of damage
Handles are free of rips or cracks
Power/data cable is free of damage
**Perform a functional test of the I I s s o o l l i i b b r r i i u u m m PE functions
**Perform system diagnostics and confirm there are no errors
*Internal components for signs of stains from fluid ingress or contamination by fully unzipping the cover
Pods do not leak or are not cracked
Hose connections to the manifold and sensor tubes (both ends) are seated
Foam is free from large tears or large gouges
Turn bladders are free of excessive wear and function
Low Air Loss fans rotate freely and are free of debris and dust
Left and right CPR releases function
Both foot box cooling fans rotate and are free of debris and dust
Pump spring mount has visible clearance from the foot box bottom
Listen for abnormal pump vibration and noise level
Clean the foot box intake metal screen
Clean the mesh on the bottom cover
Clean the fan filter screen on the foot box
Standoffs (feet of the foot box) are present
Cable tie is present on the power/data cable
N N o o t t e e
• *A zipper pull tool or equivalent is required to access the internal components of the support surface.
• **You cannot perform these preventive maintenance checks without the footboard, see the P P r r o o C C u u i i t t y y Maintenance
Manual.
• The zipper overlaps at the end and may appear to be misaligned per design.
Replace the following items every two years:
Top cover assembly
Bottom cover assembly