Stryker Isolibrium 297300000000 Manual Del Usuario página 27

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 175
D D i i s s i i n n f f e e c c t t i i n n g g
C C A A U U T T I I O O N N
• Always unplug the support surface power/data cable before you clean or disinfect.
• Do not immerse the support surface or foot box in cleaning or disinfectant solutions.
• Always disinfect the support surface between patients, to avoid the risk of cross-contamination and infection.
• Do not allow fluid to pool on the support surface. Fluids may cause degradation of components and may cause
unpredictable safety and performance of this product.
• Always wipe each product with clean water and dry after you clean or disinfect. Some cleaning and disinfecting agents
are corrosive in nature and may cause damage to the product. Failure to properly rinse and dry the product leaves a
corrosive residue on the surface of the product and may cause premature degradation of critical components. Failure to
follow these cleaning or disinfecting instructions may void your warranty.
• Do not overexpose the covers to higher concentration disinfectant solutions as these may degrade the covers.
• Do not use accelerated hydrogen peroxides or quaternaries that contain glycol ethers greater than 3% glycol ether as
they may damage the cover or the CPR strap print.
P P r r e e r r e e q q u u i i s s i i t t e e : : Minimum of two operators are required to disinfect the support surface.
Suggested disinfectants:
Quaternaries (active ingredient - ammonium chloride) that contain less than 3% glycol ether
Phenolic solution (Matar)
Chlorinated bleach solution for the Endurance cover is 10,000ppm
70% Isopropyl Alcohol
To disinfect the support surface covers after each patient use, follow these steps in order:
1. Unplug the bed from the wall outlet. Turn the battery switch to off.
2. Unplug the support surface power/data cable from the bed.
3. Remove the power/data cable from the strain relief bracket.
4. Clean and dry the support surface covers (see Cleaning (page 22)) before disinfectants are applied.
5. Apply recommended disinfectant solution with a damp cloth or pre-soaked wipes. Do not soak the support surface.
N N o o t t e e
• Follow the disinfectant manufacturer's instructions for appropriate contact time and rinse requirements.
• Raise the head of bed and foot end to help fold the head end and foot end of the support surface to disinfect the
bottom cover of the support surface.
6. Rinse the support surface covers with a clean, damp cloth to remove excess disinfectant.
7. Wipe the support surface covers with a clean, dry cloth to remove excess liquid or disinfectant.
8. Allow the support surface covers to dry before you return the product to service.
9. Turn the bed battery switch to on.
10. Plug the bed power cord into a hospital grade protective earthed outlet.
11. Plug the support surface power/data cable in through the strain relief.
N N o o t t e e - - Dry the support surface covers before you store, add linens, or place a patient on the surface.
2973-009-005 Rev AD.0
23
EN
Tabla de contenido
loading

Tabla de contenido