The TOBI HOSPI PLUS suction equipment does not need maintenance or lubrication.
It is necessary to check functioning and instrument before every use.
Unpack the instrument and always check integrity of plastic parts and feeding cable, they might have been
damaged during previous use.
Connect cable to electrical network and turn switch on.
Close the aspiration outlet with your finger and with suction regulator in maximum vacuum position check that
the vacuum indicators reaches -90 kPa (-0.90 bar) maximum.
Rotate the knob from right to left and check the aspiration regulating control.
The vacuum indicator should go down -40 kPa (-0.40 bar).
Verify that loud noises are not present, these can indicate wrong functioning.
A protection fuse (F 1 x 4A 250V) reachable from exterior and it situated in the plug protects the instrument.
For use replacing, always check the type and the range indicated.
Fault type
1. No aspiration
Jar Cap badly screwed
2. No aspiration
Lid seal not in its seat
3. The float doesn't close
If the cap has been washed,
ensure that the float is not
partially detached
4. The float doesn't close
The float it's covered by
5. Low suction
Foam inside the jar
6. No aspiration due to
flow leakage of mucus
Faults 1 - 2 - 3 - 4 - 5 - 6
None of the remedies has
achieved the desired
If the overfill security system it's activated, don't proceed with the liquid aspiration.
If the overfill security system doesn't work there are two cases:
1° case – If the overfill security system doesn't work the aspiration will be stopped by the bacteriological filter
who avoid the liquid penetration inside the device.
2° case – If both the security system doesn't work, there is the possibility that liquid comes inside the device,
in this case return the device to GIMA S.p.A. technical service.
BEFORE EVERY CHECKING OPERATION, IN CASE OF ANOMALIES OR BAD FUNCTIONING,
PLEASE CONTACT GIMA S.p.A. TECHNICAL SERVICE.
GIMA S.p.A. DOES NOT GIVE GUARANTEE IF INSTRUMENT, AFTER THE TECHNICAL
SERVICE CHECKING, APPEARS TO BE TAMPERED.
MAINTENANCE
Cause
Unscrewed the cap, then rescrew it correctly
down
Unscrew the cap and insert the seal properly
Insert the float into it's place
Unscrewed the cap, leave the and put in on
dirty material
Fill the jar to 1/3 full of ordinary water
Filter blocked
Contact the seller or GIMA S.p.A. After-
sales Assistance Service
results
12
Solution
in its seat
autoclave
Replace filter