Guidance and manufacturer's declaration – Electromagnetic Emissions
(as request by regulation EN 60601-1-2:2001)
The surgical aspirator TOBI HOSPI PLUS is intended for use in the electromagnetic environment specified
below. The customers or the user of the surgical aspirator TOBI HOSPI PLUS should assure that it's used in
Emissions Test
Irradiated / Conducted emissions
CISPR11
Irradiated / Conducted emissions
CISPR11
Harmonic emissions
IEC/EN 61000-3-2
Voltage fluctuations /
flicker emissions
IEC/EN 61000-3-3
Guidance and manufacturer's declaration – Electromagnetic Emissions
(as request by regulation EN 60601-1-2:2001)
The surgical aspirator TOBI HOSPI PLUS is intended for use in the electromagnetic environment specified
below. The customers or the user of the surgical aspirator TOBI HOSPI PLUS should assure that it's used in
Immunity Test
Electrostatic discharge (ESD)
IEC/EN 61000-4-2
Electrical fast transient / burst
IEC/EN 61000-4-4
Surge
IEC/EN 61000-4-5
Loss of voltage, brief voltage
interruptions and variations
IEC/EN 61000-4-11
Magnetic field
IEC/EN 61000-4-8
Conducted Immunity
IEC/EN 61000-4-6
Irradiated Conducted
IEC/EN 61000-4-3
Note U
is the value of the power supply voltage
T
such an environment.
Compliance
Electromagnetic environment -
Group 1
The surgical aspirator
TOBI HOSPI PLUS only used RF
energy only for its internal functioning.
Therefore its RF emissions are very low
and are not cause interference in
proximity of any Electronic appliances.
Class [B]
The surgical aspirator
TOBI HOSPI PLUS can be used in all
environments, including domestic and
Class [A]
those connected directly to the public
mains distribution that supplies power
Complies
to environments used for domestic
scopes.
such an environment.
Compliance
Electromagnetic environments -
± 6kV on contact
Floors should be wood, conceret or
ceramic tile. If floors are coverei with
± 8kV in air
synthetic material, the relative
humidity should be at least 30%.
± 2kV power supply
Mains power quality should be that of a
typical commercial environment or
hospital
Mains power quality should be that of a
± 1kV differential mode
typical commercial environment or
hospital
5%U
for 0.5 cycle
Mains power quality should be that of a
T
typical commercial environment or
40%U
for 05 cycle
T
hospital If the user of the surgical
70%U
for 25 cycle
T
aspirator TOBI HOSPI PLUS
<5%U
T
request that the appliance operates
for 5 sec
continuously, the use of a continuity
unit is recommended.
3A/m
The power frequency magnetic field
should be measured in the intended
installation location to assure that it's
sufficiently low.
3Vrms 150kHz to 80MHz
(for appliances that aren't life -
supporting)
3V/m 80MHz to 2.5 GHz
(for appliances that aren't life -
equipment)
10