User roles
2. Make the desired changes and confirm.
To remove an alarm:
1. Enable the checkbox of the alarm.
2. Select
8.5.4.5
Range settings parameter menu
Measurement range underrun, measurement range overrun and capture value
limits can be set in this menu.
The capture range refers to a range within which minor variations in measured
values (such as signal noise, variations in concentration) do not cause variations in
the display. Measured values within this range will be displayed with the capture
value. Measured values outside the capture range will be displayed with the actual
measured value.
Parameter
Overrange maxi-
mum
Overrange mini-
mum
Underrange maxi-
mum
Underrange mini-
mum
Capture enabled
Capture upper limit Capture upper limit of capture range.
Capture value
Capture lower limit Capture lower limit of capture range.
8.5.4.6
Special signals parameter menu
In this menu the user can select how special signals, warnings and faults output by
the transmitter are to be displayed by the system. Detection of dynamic transmitter
warning signals can also be activated.
To activate the detection of dynamic transmitter warning signals:
1. Select the desired warning in the Dynamic warning signal reason dropdown
list.
To add a Constant signal:
1. Select
154
to remove the alarm from the list.
Description
Range of current values for which an over range
corresponding status message is displayed on the Display
Unit for the channel. Valid measured values are displayed
up to the current value set in Overrange minimum.
Adjustable range: 20 to 24 mA;
Default setting: 20.4 mA (Overrange minimum), 21.0 mA
(Overrange maximum)
Range of current values for which a corresponding status
message is displayed on the Display Unit for the channel.
Adjustable range: 3.0 to 4.0 mA
Default setting: 3.7 mA (Underrange minimum), 3.9 mA
(Underrange maximum)
Activate the capture range to suppress interference and
variations in the display. Dräger recommends activating a
capture range in the REGARD 3000 only if a connected
transmitter does not have a capture range. Different capture
ranges in a connected transmitter and in the REGARD 3000
can lead to different measured value displays and are to be
avoided.
Quiescent value of the capture range.
.
|
Instructions for use
REGARD 3000
and a