6. Advanced use

6.1 Limited rotation
In some applications, limited space around the shaft coupling forbids the rotation of the
measuring units to the 9 or 3 o'clock position. However, it is still possible to perform the
alignment as long as the measuring units can rotate 180°.
Execute all preparation steps as per section 3.1 to 3.6.
Measuring sequence:
1. The display unit indicates that the measuring units should be placed in the 9 o'clock
position. Since you can't reach it, place the measuring units in your start position (in
our example 11 o'clock) and confirm the measurement by pressing the OK button.
2. The display unit now indicates the measuring units should be placed in the 3 o'clock
position. Rotate the measuring units by 180° (in our example to the 5 o'clock position)
and confirm the measurement: .
3. You can now complete the alignment following the instructions sequence as per
section 3.5.
6.2 Trouble shooting

6.2.1 The system does not switch on

a) Check that batteries are inserted in the right way.
b) Replace the batteries. Use Alkaline batteries for a longer battery life time.
6.2.2 No laser lines
a) Make sure that the display unit is switched ON.
b) Check the cables and connectors. Ensure that all cables are properly connected.
c) Check to see if the measuring units' warning LEDs flash.
d) Replace the batteries.
6.2.3 No measurement values
a) Check the cables and connectors.
b) Ensure that the laser lines hit the positioning detectors (see section 3.3)
c) Ensure uninterrupted travel of the laser lines.