Specifications; Warranty - Sunrise Medical DeVilbiss 9000 Serie Guía De Instrucciones

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 12
16

SPECIFICATIONS

Technical Specifications
Size .............................................................................4.0"H x 7.5"W x 10.8"D
Weight....................................................................................3.6 lbs. (1.7 kg)
Electrical Requirements (9000/9001) ...................................100-240V~, 50/60 Hz
9001 Only ........................................................................12 VDC @ 3 A max
Maximum Power Consumption ...............................................................65 watts
CPAP Pressure Range .....................................................................3-20 cm H2O
Pressure Delay Time..................................................0, 10, 20, 30 or 45 minutes
Delay Pressure Range .......................................3 cm H2O to prescription pressure
Operating Temperature Range................................41°F to 104°F (+5°C to +40°C)
Operating Humidity Range ....................................5% to 95% RH non-condensing
Operating Atmospheric Conditions .....................................Sea level to 9,000 feet
Storage & Transportation Temperature Range. .........-4°F to 140°F (-20°C to +60°C)
Storage & Transportation Humidity Range................5% to 95% RH non-condensing
Maximum Limited Pressure .................................................................28 cmH2O
Sound Level........................................30 dBA (@ 10 cmH2O, 1 meter from CPAP)
Filter specs ..................................................Standard Filter > 3 micron particles
Max flow rates
Model 9000
83 L/m
134 L/m
163 L/m
Pressure Stability
Between maintenance intervals
(maintenance interval = 1 year) Model 9001 = ± 1 cmH2O
Model 9000 Pressure swings (cmH
2
Breaths per minute
@ 7 cmH2O
@ 13 cmH2O
@ 20 cmH2O
SE-900
S p e c i f i c a t i o n s
(10.2 cm x 19.6 cm x 26.2 cm)
Optional Fine Particle Filter > 0.3 micron particles
Model 9001
73 L/m
@ 7 cmH2O (1/3 max pressure)
118 L/m
@ 13 cmH2O (2/3 max pressure)
164 L/m
@ 20 cmH2O (max pressure)
Model 9000 = ± 2 cmH2O
O peak to peak at 500mL tidal volume)
10
15
20
1.0
1.0
1.0
1.0
1.0
1.0
1.3
1.3
1.4
S p e c i f i c a t i o n s
Model 9001 Pressure swings (cmH2O peak to peak at 500mL tidal volume)
Breaths per minute
10
@ 7 cmH2O
1.0
@ 13 cmH2O
1.0
@ 20 cmH2O
1.1
Warranty.................................................................................Two-year limited
Equipment classification with respect to protection from electric shock:.......Class II
Degree of protection from electric shock:...............................Type BF Applied Part
Degree of protection against ingress of liquids:............................................None
Equipment not suitable for use in the presence of a flammable anaesthetic mixture
with air or with oxygen or nitrous oxide.
Mode of operation: ..........................................................................Continuous
Definitions
Display – purpose, type, information displayed. The LCD display shows the following
information in each model:
Model 9000
Model 9001 Description
Set Pres
Set Pres
CPAP pressure prescribed for the patient in cmH2O
Altitude
Altitude setting one of four ranges
Fine Adj
Fine adjustment for fine tuning pressure using an
external manometer
Dly Time
Dly Time
Delay time prescribed for the patient in minutes
Dly Pres
Dly Pres
Delay pressure prescribed for the patient in cmH2O
Compl Hrs
Compl Hrs
Compliance meter, shows accumulated time "while
breathing"
Unit Hrs
Unit Hrs
Unit hour meter, shows accumulated time of blower
operation
Firmware
Firmware
Firmware version installed in the CPAP
Serial #
Serial #
Serial number of the CPAP
Language
Language
Language selection for LCD display messages
Off
Off
Blower is OFF; AC or DC power is connected
Error
Error
An error has been detected by the firmware, the
CPAP is in a fail safe mode
17
15
20
1.0
1.1
1.0
1.1
1.1
1.2
SE-900
Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

Devilbiss 9001 serie

Tabla de contenido