5. Dip the tester into a solution of known temperature and allow time for the in built temperature sensor to stabilize.
6. Press
to adjust the temperature value or press the
value of the solution.
Note: To exit this program without confirming the calibration, press
Key Info
Scroll
Accept
EXIT
Temp Coefficient
1. Press
to enter setup window. Scroll down by pressing
2. Press
to select Temp Set. The display shows Set °C/°F, Temp Cal and Temp Coeff.
3. Scroll down by pressing
4. Press
to select Temp Coeff or
5. Press
to confirm the Temp Coeff value. the new value is automatically confirmed with a
Key Info
Scroll
Accept
EXIT
Calibration for Conductivity, TDS, or Salinity
For best results, periodic calibration with an accurate standard is recommended prior to measurement. Use the
calibration standard value that is close to your intended sample value.The tester will retain one calibration value in
each mode (conductivity, TDS, salinity) when the instrument is powered off. The conductivity value can be calibrated
automatically or manually, while the TDS & salinity values require manual calibration. The tester will begin in the
measurement mode that was used when it was powered off. See "Elite CTS Measurement Parameter Setting" to
change the desired parameter.
Automatic Calibration for Conductivity
1. Remove the cap and press the
2. Dip the sensor in at least 30 mm of calibration standard.
3. Stir gently and press
4. The display will show CAL followed by the default value. CAL is indicated on the display during calibration mode.
5. If the reading is within the calibration range of the automatically recognized standards; 80 (84 µS/cm),
1410 (1413 µS/cm), or 12.90 (12.88 mS/cm), the
value has been detected.
6. Press
to accept the auto conductivity standard and finish the calibration.
7. Display returns to Measurement window.
CAL
Place the tester
in the calibration
STD to start
or press
escape.
4
Measure
Set °C/°F
Settings
Temp Cal
Temp Set
Temp Coeff
Reset
to toggle between Set °C/°F, Temp Cal and Temp Coeff.
to adjust the Temp Coeff.
Measure
Set °C/°F
Settings
Temp Cal
Temp Set
Temp Coeff
Reset
to power on.
to begin the calibration.
CAL
µS/cm
844
1413
to
to confirm the calibrated value as the new temperature
.
Temp Cal
23.4°C
23.4 °C
23.4 °C
to select Temp Set.
Adjust
2.00
Temp Coeff
2.00
icon is displayed when the automatic calibration standard
COND
µS/cm
844
23.5 °C
Elite CTS Tester Operating Instructions
.