3.3 Operation with external devices

The OXYWAY Click distributor is designed so that external devices can be operated with it.
These include for example:
MEDUMAT Easy
MEDUMAT Standard
MEDUMAT Standard a
MEDUMAT Basic p
OXYMAND demand valve
OMNIVAC secretion suction
Caution!
If external devices are used, pay particular attention that the total flow
taken off does not exceed the maximum flow of the pressure reducer.
If all inhalation outputs and an external device are used simultaneously,
oxygen output may be slightly reduced (see "9.2 Quantity of oxygen sup-
plied by the OXYWAY Click distributor" on Page 16).