2.2.10 Discharge valve installation
Note: Make sure that the discharge valve is open before installation, in case of damage of the reactor body.
Please refer to Figure 2.2.10.
Fit upper part of the flange (59) and a lid gasket (60) to the glass drain port (9) and the lower part of the flange and gasket
to the discharge valve (14). Put sealing ring (61) between the upper and lower flange, keep them in the right position and
tighten the butterfly nut.
2.2.11
Vacuum gauge and control box installation
Please see chapter 1.4 to find the installation position of Vacuum gauge (4) and control box (17).
2.2.12
Vessel jacket circulating pipe kit installation
The lower connecting joint of the vessel jacket is inlet, the upper is outlet. Please refer to figure 2.2.10.2 to install it.
48
Figure 2.2.10 RE‐650 discharge valve installation
Figure
2.2.12
9 Reactor Vessel
14 Discharge valve
59 Flange
60 Lid gasket
61 Sealing ring
62 Connector
63 Ball valve
64 Three‐way adapter
65 Hose bracket
66 Flat washer
67 Stainless steel hose
68 Connection elbow
Circulating pipe kit installation