Cal Compensation
Standard
1.25
µS/cm
25.0
°C
Conductivity Calibration
Type = 1 point
1.25
µS/cm
25.0
°C
A Point1 = 0.055 µS/cm
C = 0.057 µS/cm m
A
1.25
µS/cm
25.0
°C
C M=0.1000
A=0.0000
Save Calibration? Yes
12
Choose the compensation mode by using the
Press [ENTER] to confirm the compensation mode. .
m
Select "1 point" instead of "2 point Calibration" by pressing the
by the [ENTER] key.
m
CAUTION: Rinse sensors with a high-purity water solution before every calib-
ration to prevent contamination of the reference solutions.
Place the electrode in the reference solution.
Enter the value of Point 1 and press the [ENTER] key.
After the calibration of the cell multiplier or slope calibration factor "M" i.e. cell constant
and the Adder or offset calibration factor "A" are displayed.
Select "Yes" to save the calibration values and the successful Calibration is confirmed
m
on the display.
The user gets the message "Re-install sensor" and "Press [ENTER]" on the display.
After pressing [ENTER] the M200 easy returns to the measuring mode.