Lovibond Tintometer XD 7500 Guia De Inicio Rapido página 12

Ocultar thumbs Ver también para Tintometer XD 7500:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 35
EN
English
Switching ON
Self test start
Self test
12
4
Operation
4.1
Switching the photometer ON/OFF
Self test
Please make sure no cell is inserted and the cover is closed.
Then press <START/ENTER>
Setup
Self test
Please make sure no cell is inserted and the cover is closed.
Then press <START/ENTER>
All cells must be removed and the cell shaft covers must be closed during the
self-test
Self test
System test
System test
Filter test
Lamp test
Wavelength calibration
You can review and print the result of the self-test with the [Info]
function key.
1 Use <ON/OFF> to switch the
photometer ON.
The display shows
16.01.15 9:52
– the dialog Self test (with
deactivated user management)
or
– the dialog Login
(with activated user
management).
Info
2 Remove all cells and close the cell
16.01.15 9:52
shaft cover.
3 Use <START·ENTER> to start the
self-test.
The photometer carries out the
self-test.
The self-test includes:
16.01.15 9:52
– the test of memory, processor,
internal interfaces,
filter and lamp
– a calibration for each shaft
length
After the self-test is finished, the
main menu will be displayed on
the screen.
XD 7500
ba77195y01
09/2017
Tabla de contenido
loading

Tabla de contenido