Clean The Meter; Apply Silicone Oil To A Sample Cell - Hach 2100Q Manual Básico Del Usuario

Ocultar thumbs Ver también para 2100Q:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 35
Option
Description
Signal
The Signal Average mode compensates for reading
Average
fluctuations caused by drifting of sample particles
through the light path.
The X-bar icon is shown on the display when signal
averaging is on.
The Signal Average mode measures 12 times and
starts to show the average after three readings. The
final result is the average of all 12 readings.
Rapidly
The Rapidly Settling Turbidity (RST) mode
Settling
calculates and continuously updates the turbidity
Turbidity
reading of the sample to a confidence of 95%,
(RST)
based on the accumulated trend of the real time
measured values.
The RST mode is best used on samples that settle
rapidly and continuously change in value. The
reading is based on a correctly prepared sample
that is homogeneous at the beginning of the
reading. It is best applied to samples that are
greater than 20 NTU. The sample must be mixed
thoroughly by inversion immediately before inserting
it into the meter.
The target icon is shown on the display when the
Rapidly Settling Turbidity is on.
The Rapidly Settling Turbidity reads and calculates
five readings while showing intermediate results.
Section 8 Maintenance
C A U T I O N
Multiple hazards. Only qualified personnel must conduct the
tasks described in this section of the document.
12 English

8.1 Clean the meter

The meter is designed to be maintenance-free and does not require
regular cleaning for normal operation. Exterior surfaces of the meter may
be cleaned as necessary.
Note: Do not clean the meter with solvents to avoid damaging the material.
1. Clean the meter with a dust- and lint-free dry or slightly damp cloth. A
mild soap solution can also be used for liposoluble contamination.

8.2 Apply silicone oil to a sample cell

Sample cells and caps must be extremely clean and free from significant
scratches. Apply a thin coating of silicone oil on the outside of the
sample cells to mask minor imperfections and scratches that may
contribute to light scattering.
Note: Use only the provided silicone oil. This silicone oil has the same refractive
index as the sample cell glass.
Tabla de contenido
loading

Este manual también es adecuado para:

2100qis

Tabla de contenido