Error Handling - HEBU medical HRF 100 x-touch Instrucciones Para El Uso

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

HBS x-touch 200i

7.12 Error handling

In the event of an error, a message appears on the display.
The following error messages can occur:
Error
No. Cause of failure
1
"Power timeout error"
Err 1
Total Failure
10
"Power Error"
2
"NE electrode error"
Err 2
Mono - Cut hand /FS1
6
Modules
Error"
Failure
"Mono – Coag (Bi) hand
7
/FS1 Error"
4
"Power Supply 1 Error"
Err 03
PS Failure
5
"Power Supply 2 Error"
11
"EEROR in CPU ROM"
Err 5
CPU Failure
12
"EEROR in CPU RAM"
If a different numerical code is displayed or the device has other recognizable defects, please contact
Error management
The supply voltage was not reached. Check the
controller and the power pack, replace if necessary
The difference between the display and the measured
values of the monitoring is greater than 20%. Check
the controller, the interface and the power pack,
replace if necessary.
Check the connection on the device and the electrode
on the patient, if necessary replace the NE and / or
connection cable.
Check CUT connections, foot switch settings, check
settings.
Check COAG connections, foot switch settings, check
settings.
The block voltage required to correct the power factor
is missing. Check the power connection, check the
power supply unit, replace if necessary.
.
There is no voltage required from the block for
regulating the power voltage for the power generator
when idling. Check the power connection, check the
power supply unit, replace if necessary..
Incorrect check sum of the ROM memory from the
control processor. Check controller, replace if
necessary
Incorrect check sum of the RAM memory from the
control processor. Check controller, replace if
necessary.
EN
53
Tabla de contenido
loading

Tabla de contenido