3. Operation
3.1 Preparing for use
3.2 After use
1.
10
EN
Operation
1. Make sure that the dial head of the flowmeter is in po-
sition "0".
2. Connect the OXYWAY Click to the central gas system.
This is done by pushing the inlet connector of the
OXYWAY Click into the socket of the central gas sys-
tem until you hear it snap into place.
3. Connect the inhalation accessories to the OXYWAY
Click.
4. Use the dial head to set the desired flow.
5. The OXYWAY Click is now ready for use.
1. Set the dial head to "0". This is necessary to prevent
pressure hammer when the oxygen supply is opened
again later.
Important
OXYWAY Click is not a shutoff valve! Small amounts
of oxygen can still escape even when it is completely
closed.
2. If the OXYWAY Click is no longer needed here,
remove it from the supply socket.
2.