Implementation of the provision Total suspended solids
Select the method on the device
Homogenize ml of the water sample in a blender on high speed for minutes
10 ml
Fill 24 mm vial with 10 ml
deionised water .
Zero
Press the ZERO button.
Mix homogenised water
sample thoroughly.
EN   Method Reference Book   1.0
Close vial(s).
Remove the vial from the
sample chamber.
24 mm
Pre-rinse vial with water
sample.
Suspended solids 24 / M384
Sample
Place sample vial in the
sample chamber. • Pay
attention to the positioning.
10 ml
24 mm
Fill 24 mm vial with 10 ml
prepared sample .
EN
7