CU-EMA+
Status text
Possible cause
H2o.1
• Moisture
alarm mois-
ture detector 1
init
• Initialisation
phase
PuMP • Pumps deac-
tivated
dt
Active Delta T con-
trol only: The block
temperature is out-
side the defined
temperature
range.
• Cooler is still in
the "break-in
phase"
• Fluctuating
ambient tem-
perature
• Insufficient
cooling capa-
city
• Excess/low
temperature
(Flashing)
BX440029 ◦ 10/2020
Action
• Dry
• Check condensate
trap
• Wait
• Reactive pumps via
menu
• Wait to see if the
target temperature
will be reached
• Check the ambient
temperature /
power supplied
• Process-related: Ad-
just alarm limits
• see chapter
"Troubleshooting"